Calibration Station for the GX-6000

SDM-6000-2

Contact Our
Sales

The SDM-6000 is a versatile maintenance tool that automates the service of a GX-6000 portable gas monitor. This docking station can be used as a stand alone device for charging, calibration and bump testing, with all records stored automatically for future upload to a PC via SD card (PC software and SD card included).

Modular Design 
SDM-6000 calibration stations are modular by design and can easily connect to each other without the need of any tubes or cables. One power source and one set of calibration equipment can provide service for up to 10 SDM-6000 calibration stations.

Up to 10 calibration stations can also be connected to one PC for automated charging, calibration, bump testing and archiving of data. Even if instruments are turned off, automated routines will power up the instruments and run as scheduled. All data collected from the PC Controlled software can also be centralized in one network location accessible by anyone with the PC Controlled software with access to the network location.



  • Stand-alone station (no PC Required)
  • Charge, calibrate, bump test, and alarm test
  • Included software collects calibration, bump test, alarm history, and exposure date from instruments
  • Automatically power instruments on / off for scheduled calibration
  • Accepts GX-6000 with or without rubber boot
  • Module Design

    • Simultaneously bump / calibrate, charge, and collect data up to 10 instruments
    • No tubing, adaptors or wires needed to connect
    • One power source for all stations
    • Configure GX-6000 parameters
    • Centralize instrument data from multiple stations
      and locations to one network location
    • Ethernet port for LAN connectivity
    • Schedule automatic calibration / bump test
    • Optional valve expansion module for up to six gas cylinders
    • Optional battery operated printer
SPECIFICATIONS
DisplayTouch panel TFT liquid crystal (3.5 inch, 320×240 image element QVGA, RGB1677 million colors) Back light
LED3 color [Green, Red, Orange]. Each status is indicated by Lighting / Blinking
BUMP testStandard setting values

AIR Flush time: 15sec, GAS time: 20sec, AIR Purge time: 5sec

Threshold value: ±50%, Auto Calibration: OFF, Auto alarm function check: ON

Auto print: OFF

Setting range

AIR Flush time: 15~120sec

GAS time: 20~120sec

AIR Purge time: 5~120sec

Threshold value: ±10~50%

Auto calibration: ON, OFF

Auto alarm function check: ON, OFF

Auto print: ON, OFF

Bump test result: Displayed on TFT liquid crystal

CalibrationStandard setting values

AIR Flush time: 30sec, GAS time: 90sec, AIR Purge time: 30sec

Auto alarm function check: ON, Auto print: OFF

Setting ranges

AIR Flush time: 15~120sec

GAS time: 20~120sec

AIR Purge time: 5~120sec

Auto alarm function check: ON, OFF

Auto print: ON, OFF

Calibration result: Displayed on touch screen display

Power (AC adaptor)100~240VAC~50/60Hz
Dimension, weight5.11”(W) x 13’(H) x 7.2”(D)in Approx. 3.7lb
Operation temperature, humidity0~104F, below 95~RH (no condensing)
USB printer portA type: Connects SDM-6000 printer (SII, DPU-S245) via USB cable
RJ-45 connectorEthernet port for a LAN connection
SD card slotCompatible with SD & SDHC cards
Joint connectorConnecting SDM-6000 (Max 10 units) and SV-3
Inlet4 inlets: GAS 1 / GAS 2 / GAS 3 / AIR
CertificationCE marking
Standard FunctionsCharging, Cal. cap fitting status monitoring, Power ON function, Alarm activation check, multi language (English, German, Japanese, Korean, Russian, Portuguese, Spanish, Italian, French), backlight adjusting, buzzer function, self-diagnosis for instruments connection
Standard accessoryOperating manual, filter, connecting plate, wall mount plate, PC controller software
SoftwarePC controller software included as standard. Windows 7 & 10 compatible
Optional accessorySV-3 solenoid valve adds 3 additional calibration gas inputs

AC adaptor (15W output), AC adaptor (100W output)

Calibration Station utility software

Calibration gas & regulators

Tubing

Battery powered printer

Specifications subject to change without notice.

Ordering Information

The SDM-6000 can be ordered using the part numbers from the list below.

Docking Station

81-SDM6000GX-6000 docking station with particle filter, connection brackets, SD Card, & USB SD card reader.
81-SDM6000-01Calibration station for SDM-6000, includes particle filter, connection brackets, exhaust tubing, USB cable for PC, SD card, USB SD card reader, ethernet cable, and single unit AC adapter
81-SDM6000-02Calibration station for SDM-6000, includes particle filter, connection brackets, exhaust tubing, USB cable for PC, SD card, USB SD card reader, ethernet cable, and multi-unit AC adapter
81-SDM6000-03Calibration station for SDM-6000, includes particle filter, connection brackets, exhaust tubing, USB cable for PC, SD card, USB SD card reader, ethernet cable, single unit AC adapter, demand flow regulator, and cal gas tubing
81-SDM6000-04Calibration station for SDM-6000, includes particle filter, connection brackets, exhaust tubing, USB cable for PC, SD card, USB SD card reader, ethernet cable, single unit AC adapter, demand flow regulator, cal gas tubing, 58AL cyl H2S/CO/CH4/O2, and cylinder holder
81-SDM6000-05Calibration station for SDM-6000, includes particle filter, connection brackets, exhaust tubing, USB cable for PC, SD card, USB SD card reader, ethernet cable, single unit AC adapter, demand flow regulator, cal gas tubing, 58AL cyl 100 ppm isobutylene/H2S/CO/CH4/O2, and cylinder holder
81-SDM6000-06Calibration station for SDM-6000, includes particle filter, connection brackets, exhaust tubing, USB cable for PC, SD card, USB SD card reader, ethernet cable, single unit AC adapter, demand flow regulator, cal gas tubing, 34AL cyl 100 ppm isobutylene/air, and cylinder holder
81-SDM6000-07SDM-6000 calibration station for GX-6000, includes particle filter, connection brackets, exhaust tubing, USB cable for PC, SD card, USB SD card reader, ethernet cable, single unit AC adapter, demand flow regulator, cal gas tubing, 34AL cyl H2S/CO/CH4/O2, and cylinder holder
81-SDM6000-08SDM-6000 calibration station for GX-6000, includes particle filter, connection brackets, exhaust tubing, USB cable for PC, SD card, USB SD card reader, ethernet cable, single unit AC adapter, 2 demand flow regulators, cal gas tubing, 34AL cyl H2S/CO/CH4/O2, 34AL HCN/N2, and 2 cylinder holders
81-SV3SV-3 solenoid valve unit, with connection brackets, for 6 cylinder expansion with SDM-6000
Call for additional configurations